Beamlines SER-CAT operates two insertion-device beamlines at Sector 22 of the Advanced Photon Source for macromolecular crystallography, with on-site and remote data collection. Beamline 22-ID-D: Variable Wavelength Operations 22-ID-D Technical Specifications X-ray Detection System: Primary Detector: Dectris Eiger X16M Fluorescence Detection: Amp-Tek 123 SDD X-ray Spectrometer Minimum Crystal-to-Detector Distance: 150 mm Energy and Wavelength Capabilities: Energy Range: 6.0 - 24.0 keV Wavelength Range: 2.1 - 0.52 Ångström (approximate) Configuration: Variable wavelength operations for multi-wavelength anomalous diffraction (MAD) experiments Goniometer and Sample Handling: Diffractometer: MD3 vertical omega-axis system (Arinax) Sample Mounting: Automated crystal mounting system compatible with MD3 platform Sample Capacity: 20 Unipuck positions with complete pin accessibility Additional 10 Unipuck positions with partial accessibility (pins 6, 7, 16 restricted) Beam Characteristics: Flux: 7 × 10¹² photons/second Beam Size: 10 μm × 10 μm (FWHM) Control Software: SERGUI beamline control and data collection platform Beamline 22-ID-E: High-Throughput Fixed Wavelength Technical Specifications X-ray Detection System: Primary Detector: Rayonix MX300-HS high-speed detector Fluorescence Detection: Amp-Tek FAST 3DD XR-100 SDD X-ray Detector Minimum Crystal-to-Detector Distance: 150 mm Energy and Wavelength Capabilities: Fixed Energy: 12.700 keV Fixed Wavelength: 0.97 Ångström (approximate) Configuration: Optimized for high-throughput native data collection Goniometer and Sample Handling: Diffractometer: MD2 horizontal omega-axis system (Arinax) Sample Mounting: Automated crystal mounting system compatible with MD2 platform Sample Capacity: 6 Unipuck positions with complete pin accessibility Mounting Efficiency: 30-second dismount/mount cycle time Beam Characteristics: Flux: 7 × 10¹² photons/second Beam Size: 10 μm × 10 μm (FWHM) Control Software: SERGUI beamline control and data collection platform Comparative Beamline Parameters Specification 22-ID-D 22-ID-E Primary Application Variable wavelength MAD/SAD High-throughput screening Energy Range 6,000 - 24,000 eV 12,700 eV (fixed) Detector Technology Dectris Eiger X16M Rayonix MX300-HS Beam Flux 7 × 10¹² ph/s 7 × 10¹² ph/s Beam Size (FWHM) 10 μm × 10 μm 10 μm × 10 μm Minimum Detector Distance 150 mm 150 mm Sample Capacity (full access) 20 Unipuck positions 6 Unipuck positions Diffractometer Orientation Vertical omega-axis Horizontal omega-axis Capabilities (Both Beamlines) Microbeam and crystal rastering Automated sample mounting Automated data reduction, including cloud-based AXDPRO auto-processing On-site cluster for structure determination Wet lab with crystallization and crystal-freezing facilities software and Data Processing HKL-2000, XDS, DIALS, KYLIN, AXDPRO Phasing & refinement PHENIX, CCP4, SHELXD, Phaser, SOLVE/RESOLVE, AlphaFold2 Model building Coot, ARP/wARP Data transfer & storage Globus, 450 TB storage, USB drives, 10 Gb network See the Remote Data Collection(SERGUI).